Model 106L臭氧分析儀設計用來地測量大氣中的臭氧濃度 在一個大的動態范圍內從測量由下限1.5ppb到100ppmv的臭氧含量 三擋量程0-1,1-10,10-100ppm,自動換擋。 它使用了基于254nm的紫外吸收技術。 精度:1.5 ppbv或2%,取大者。 準確度:1.5 ppbv或2%,取大者。 尺寸:10.1×18.8×21.6cm。 重量:1.9Kg 電源要求:11-14 V DC,通常在12V時 電流約為300mA, 功耗約為3.6W
For industrial ozone applications, 2B Tech offers the Model 106-L, 106-M and 106-H Ozone Monitors, where L, M and H refer to Low (0-100 ppm), Medium (0-1000 ppm) and High (0-20 wt%) ozone concentrations, respectively. These instruments have different optical path lengths in order to accommodate this extremely wide range of ozone concentrations, spanning more than eight orders of magnitude. Additionally, the Model 106-H was designed with a flow through path that can be pressurized to up to 50 psi for in-line measurements with ozone generators. All three models are available in either a standard enclosure or NEMA enclosure.
The Model 106 series was designed as an "ozone monitor on a board" in which nearly all fo the components are mounted directly to the printed circuit board. Thus, the Model 106 may be purchased without the enclosure for OEM applications. See: Model OEM-106.
Specifications for Different Optical Benches
Model
Range
Resolution
Precision & Accuracy
106-L
0-100 ppm
0.0001 ppm (0.1 ppb)
Higher of 2 ppb or 2% of reading
106-M
0-1000 ppm
0.01 ppm
Higher of 0.01 ppm or 2% of reading
106-H
0-20 wt%
0.01 wt%
Higher of 0.01 wt% or 2% of reading
General Specifications
Measurement Principle
UV Absorption at 254 nm
Measurement Interval
10 s for Models 106-L & 106-M, 2 s for Model 106-H
Data Averaging Options
10 s, 1 min, 5 min, 1 hr
Flow Rate
~1 Liter/min; Up to 20 L/min for Model 106-H
Data Storage
32,736 lines (10 s avg. = 3.7 days; 5 min avg = 113 days)
Choice of Units
ppb, ppm, pphm, µg m-3, mg m-3 (vol%, wt% air, wt% O2 for Model 106H)