OVERVOLTAGEPROTECTION(OVP)
SETTINGACCURACY:±(0.25%+0.25V).
RESOLUTION:125mV.
RESPONSETIME:<1.5ms.
OVERCURRENTPROTECTION(OCP)
SETTINGACCURACY:±(0.25%+0.10A).
RESOLUTION:25mA.
RESPONSETIME:<1.5ms.
OVERTEMPERATUREPROTECTION(OTP)
OUTPUTTURN-OFFTEMPERATURE:>93°C(typical).
RESPONSETIME:<1.5ms(typical)8.
NOTES
1.Specificationsbasedonusingremotesenseconnections.For2-wireconnections,addanoffsetof0.5mV/A
(frontterminals).
2.6½-digitresolution,1PLCreadingrate,Filteron,Autozeroon.
3.100V/sto1000V/sslewrateislimitedto5Vchangesatamaximumof3Aforthe2281S-20-6supply.
4.6½-digitresolution,1PLCreadingrate,filteron,autozeroon.
5.Settings:Autozerooff,0.002PLC,ArmSource:external,TriggerSource:Immediate.
6.Timeincludestriggerdetection,latencyplusjitterofstartofmeasurement,andmeasurementintegrationtime,
0.002PLC.
7.Settings:AutozeroOff,OutputOn,OutputDelayOff,FixedSource:Voltage.
8.Settings:AutozeroOff,OutputOn,OutputDelayOff,FixedSource:Voltage,ArmorMeasurecount1000.